Article
Comparison of Independent Optical Frequency Measurements Using a Portable Iodine-Stabilized Nd:YAG Laser 公开 Deposited
- Abstract
A comparison of two independent absolute optical frequency measurements has been carried out between the National Metrology Institute of Japan/National Institute of Advanced Industrial Science and Technology (NMIJ/AIST), Tsukuba, Japan and the JILA (formerly the Joint Institute for Laboratory Astrophysics), Boulder, CO, using a portable iodine-stabilized Nd:YAG laser. The agreement between the two absolute measurements is better than the measurement uncertainty of 6.7×10-13 that can be attributed to the reproducibility limitations of the portable laser. This comparison is used to confirm the measured absolute frequency of an iodine-stabilized Nd:YAG laser at NMIJ/AIST (Y3), which is reported to the Consultative Committee for Length (CCL) for the determination of the absolute frequency value of iodine-stabilized Nd:YAG lasers.
- Creator
- Date Issued
- 2003-04-01
- Academic Affiliation
- Journal Title
- Journal Issue/Number
- 2
- Journal Volume
- 52
- File Extent
- 240-244
- 最新修改
- 2020-01-14
- Resource Type
- 权利声明
- DOI
- 10.1109/TIM.2003.811676
- ISSN
- 0018-9456
- Language
关联
单件
缩略图 | 标题 | 上传日期 | 公开度 | 行动 |
---|---|---|---|---|
|
comparisonOfIndependentOpticalFrequencyMeasurementsUsingA.pdf | 2019-12-05 | 公开 | 下载 |
|
comparisonOfIndependentOpticalFrequencyMeasurementsUsingA.pdf | 2019-12-05 | 公开 | 下载 |